Calibration must be performed under controlled laboratory conditions using a specialized parallel-plate charging system. Do not attempt to adjust internal pots manually, as this voids the manufacturer's warranty.
To maintain the ±10% accuracy tolerance required by international ESD standards, the FMX-002 must be calibrated once a year. Calibration requires a specialized static test fixture and a precision high-voltage power supply.
The face of the FMX-002 features an LCD screen and three multi-function buttons:
Turn on the device away from any known static fields by pressing the red button. Ensure that the battery icon on the LCD does not indicate a low charge. Step 2: Grounding the Device simco fmx002 manual full
Offers reliable readings with an accuracy of approximately ±10% of the reading.
For those who haven't yet absorbed the manual, here is a concise breakdown. The SIMCO FMX-002 is a handheld electrostatic field meter used to measure the surface voltage of an object without making physical contact.
: Press the HOLD button to freeze a reading on the display, which is helpful in tight spaces where the screen is hard to see. Technical Specifications SIMCO FMX-004 Electrostatic Field Meter User Manual Calibration requires a specialized static test fixture and
The FMX-002 provides accurate readings within a compact, ergonomic frame. (low range) and up to (high range). Accuracy: of the reading. Measuring Distance: (approx. 1 inch). Ion Balance Range: (requires optional balance plate). Response Time: Less than 1 second. Power Source: 9V manganese battery (6F22). Weight: ) including battery. Operating Instructions
Displays remaining battery capacity. Replace the 9V battery immediately when the low battery icon flashes to prevent erroneous readings. HOLD: Appears when the freeze-frame feature is active.
To ensure accurate, repeatable data, follow these step-by-step operating procedures: Step 1: Grounding the Unit Step 2: Grounding the Device Offers reliable readings
By following this guide, you'll be well on your way to becoming proficient in using the Simco FMX002 and taking full advantage of its features and capabilities.
range, a 5-minute auto-power-off, and an ion balance mode for measuring ionizers up to . For the full instructional overview, visit Simco Japan . SIMCO Japan - FMX002 Fieldmeter
| Symptom | Most Likely Cause (per manual) | Solution | | :--- | :--- | :--- | | Device won't power on | Dead battery or reversed polarity | Replace 9V battery; check orientation | | Erratic, non-zero reading in air | Dirty sensor grille | Gently blow out with clean, dry compressed air | | Drifting measurement | High humidity or nearby ionizer | Test in a different environment (manual suggests <70% RH) | | Display dim or flickering | Low battery | Replace battery immediately to prevent false readings | | No response to static charge | Internal sensor failure | Return to SIMCO for service (do not open the unit) |
), precisely guided by two LED "ring lights" that must overlap on the target surface. Response Time : Less than with a display renewal rate of Construction
Many users search for the manual because they press "Hold" and think the unit is broken.