and Clinical Reference Manual for complete system information. or a guide for NIBP manifold replacement 04-GE-E-PSMP-01-Carescape-Module-Manual.pdf
Connect a variable AC supply (Variac) to the input. Slowly ramp the input voltage up to 110V/220V while monitoring the input current. Verify that the status LEDs illuminate and outputs read a stable +24V DC and +5V DC.
E-PSMP-01, PATIENT SIDE MODULE, ENGLISH, Patient Monitoring | GE HealthCare Service Shop India. GE HealthCare
Utilizes an LLX resonant or forward-switching topology driven by high-speed PWM controller ICs. Switching transistors (MOSFETs) chop the high-voltage DC into high-frequency square waves fed into an isolation step-down transformer. e-psmp-01 service manual
Remove the four external chassis screws securing the outer metal protective cover. Lift the cover straight up.
Calibration is critical for medical devices, directly impacting the accuracy of diagnoses. This section provides step-by-step instructions for calibrating:
Planned maintenance kits for 1-year or 4-year cycles are recommended to maintain operational longevity. Go to product viewer dialog for this item. E-Psmp-01 Module Bus Connector Board Verify that the status LEDs illuminate and outputs
This comprehensive technical reference details everything needed to troubleshoot, repair, and maintain this widely deployed vital signs module. Core Specifications & Monitored Parameters
While the full procedure is lengthy, it provides a logical sequence, starting with removing the front cover:
: Non-invasive blood pressure using the oscillometric method. : Pulse oximetry for blood oxygen saturation. Invasive Pressure (IBP) it provides a logical sequence
High-speed Schottky barrier diodes rectify the high-frequency AC from the transformer. An LC filtering network smooths the output. Linear optocouplers isolate the secondary voltage feedback loop from the primary PWM controller to maintain tight voltage regulation. 4. Preventative Maintenance and Inspection Schedule
Provides dual-channel processing for arterial or venous pressure transducers alongside dual resistance-temperature probes. Preventative Maintenance & Planned Calibration
The final sections of the manual cover the administrative side of servicing the E-PSMP-01, including ordering parts and ensuring regulatory compliance.
Adjusting settings to ensure the host monitor recognizes the parameters correctly. C. Maintenance and Calibration